Model Number CM2000
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during cardiomems implant while attempting to deploy the sensor, the proximal loop of the sensor stayed attached to the device sheath.After unsuccessful troubleshooting, which included multiple breath holds and inserting a swan catheter next to the delivery catheter to try to ¿bump¿ the device off, the sensor needed to be snared in order to extract from the patient, and the implant procedure was aborted.An additional access site was used to remove the device.The physician confirmed that this issue caused a clinically significant delay as the procedure took longer than planned due to length of troubleshooting and the intervention needed to remove the device.A second attempt at implant is not scheduled at this time.
 
Additional Manufacturer Narrative
The investigation codes along with investigation results will be provided in a subsequent submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF Sensor Delivery System
Common Device Name
System, Hemodynamic, Implantable
Manufacturer (Section D)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer (Section G)
Abbott Medical
387 nerem st nw suite 500
atlanta GA 30313
Manufacturer Contact
fredric gold
168 middlesex turnpike
burlington, MA 01803
+1(651)756-5400
MDR Report Key50869096
Report Number3004936110-2026-01579
Device Sequence Number15781370
Product Code MOM
UDI-Device Identifier05414734509237
UDI-Public(01)05414734509237(10)11234733(17)280209
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P100045
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCM2000
Device Catalogue NumberCM2000
Device Lot Number11234733
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/09/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexFemale
Patient Weight310 LB
-
-