| Catalog Number |
00770301500 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that skingraft mesher doesn't make smooth holes for grafts.There is no information provided regarding the timing of the event or if there was any patient involvement/harm that may have occurred.During device evaluation, it was discovered that the cutter failed the sample mesh test.Due diligence is complete as multiple attempts were made; however, no further information is available.As no additional information is available, we are unable to provide further information.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the device failed the sample mesh test during evaluation.The cutter was damaged; however, the cutter is not repairable.The cutter was returned as is.Review of the device history record identified no deviations or anomalies during manufacturing.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2: foreign: finland.
|
| |
|