| Model Number |
12773-03 |
| Medical Device Problem Codes |
Material Separation (1562); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that suture placement in the right common femoral artery using a prostyle device using the pre-close technique via a 6f sheath hole prior to a transcatheter aortic valve replacement (tavr) interventional procedure.The sutures of two prostyle devices were successfully placed.The sheath was upsized to a 16f sheath, and the tavr procedure was completed.Reportedly post procedure, during knot advancement of the second device suture, a frayed suture was noted.The suture break ultimately occurred after the frayed suture while tightening the knot.The first pre-placed suture worked as intended and achieved hemostasis.There was no reported adverse patient effect.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the reported information, the reported suture separation and frayed suture could not be determined.Factors that may contribute to suture separation during knot advancement include, but are not limited to, user technique (tensioning angle not coaxial) or suture/ nick damage.Factors that may contribute to frayed suture include, but are not limited to, an interaction of the device with patient anatomy where the suture got pushed over the posterior side of the guide tube edge as it tract during deployment or suture snag during harvest.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|
|
|