• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; MONITOR,BP,AUTO,DIGITAL,ADULT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; MONITOR,BP,AUTO,DIGITAL,ADULT Back to Search Results
Catalog Number MDS3001
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "there air leak in the hose near cuff.She found a hole where the hose inserts into the cuff.".
 
Additional Manufacturer Narrative
The customer reported an air leak in the hose near cuff and found a hole where the hose inserts into the cuff.A sample/photo was provided, and the investigation stated that medline received the bp monitor in used condition.The investigation results were unconfirmed, and the complaint was not confirmed.No confirmed root cause was identified; the investigation stated the unit was out of warranty and that because the sample is over 2 years old, this is most likely the reason why the sample has started to show signs of malfunction and wear/tear, and the division will continue to monitor the issue for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
MONITOR,BP,AUTO,DIGITAL,ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
ankitha komirishetty
three lakes drive
northfield, IL 60093--275
MDR Report Key50869423
Report Number1417592-2026-48849
Device Sequence Number10715150
Product Code DXN
UDI-Device Identifier10080196525140
UDI-Public10080196525140
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K183534
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS3001
Device Lot Number43321080004
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/05/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2024
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-