• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Arthrex, Inc. SUTURELASSO, 45 DEGREE CURVE RIGHT; ORTHOPEDIC MANUAL SURG INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Arthrex, Inc. SUTURELASSO, 45 DEGREE CURVE RIGHT; ORTHOPEDIC MANUAL SURG INSTR Back to Search Results
Model Number SUTURELASSO, 45 DEGREE CURVE RIGHT
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/05/2026
Type of Reportable Event Malfunction
Event or Problem Description
On 08-sep-2026, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-4068-45r 45° curve right suturelasso tip became bent when using the instrument for the second anchor.The surgeon attempted to straighten it using a kelly clamp, and when reintroducing the device to pass through the labrum, the tip broke.This was discovered during a shoulder instability repair with knotless pushlock procedure with no patient harm.No fragments remained in the patient.There was no delay to the procedure, and no additional anesthesia was required.The case was completed with another suturelasso.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTURELASSO, 45 DEGREE CURVE RIGHT
Common Device Name
ORTHOPEDIC MANUAL SURG INSTR
Manufacturer (Section D)
Arthrex, Inc.
1370 creekside boulevard
naples FL 34108 -194
Manufacturer (Section G)
Arthrex, Inc.
1370 creekside boulevard
naples FL 34108 -194
Manufacturer Contact
jared engle
+1(800)933-7001
MDR Report Key50869611
Report Number1220246-2026-04869
Device Sequence Number9425953
Product Code LXH
UDI-Device Identifier00888867031494
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTURELASSO, 45 DEGREE CURVE RIGHT
Device Catalogue NumberAR-4068-45R
Device Lot Number5027156438
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-