| Model Number |
SUTURELASSO, 45 DEGREE CURVE RIGHT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 08-sep-2026, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-4068-45r 45° curve right suturelasso tip became bent when using the instrument for the second anchor.The surgeon attempted to straighten it using a kelly clamp, and when reintroducing the device to pass through the labrum, the tip broke.This was discovered during a shoulder instability repair with knotless pushlock procedure with no patient harm.No fragments remained in the patient.There was no delay to the procedure, and no additional anesthesia was required.The case was completed with another suturelasso.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|