The customer reports the blood pressure gauge doesn't work at all and cuff does inflate.Medline received 1 handheld pvc aneroid; upon visual inspection there was no cosmetic damage, and functional testing was performed on the cuff by inflating it to 300mmhg for 10 sec, repeated 5 times, and 5 out 5 times it held the pressure.The complaint was not confirmed, and the cause was documented as not confirmed with received sample.No further action was taken because objectionable trends were not seen for this issue, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|