| Model Number |
B35300 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that bilateral elevated impedance occurred during implantable neurostimulator (ins) to lead connection on contacts 3 and 11.Patient was asleep during procedure and no further intervention planned at this time. connection at head was opened, clean and dried bilaterally with no change in impedance.Connection at ins was opened cleaned and dried with no change in impedance.This process was repeated with no change.New extension was opened and tested on both sides outside of the body with no change in impedance.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|