The customer reported receiving inaccurate readings on their meter for the control testing.Medline received 1 evencare proview blood glucose meter; the sample came with batteries, a control cr2032 battery was inserted, and the sample was tested according to the instructions for use.Test strips and control solutions were not sent for review, so other test strips and control solutions were used for testing, with results of level 2: 69 and level 3: 219.The complaint was not confirmed; no indications of power or display issues were experienced, the sample turned on and displayed readings within the test settings, and the sample worked as intended.Cause was not confirmed with received sample, and based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|