| Model Number |
12773-03 |
| Medical Device Problem Codes |
Failure to Cycle (1142); Unintended System Motion (1430); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this was venous closure of the right common femoral vein using a prostyle device using the pre-close technique via a 10f sheath hole prior to an ablation interventional procedure.Reportedly, during step 2, the plunger was pulled back/dislodged (protruding) but still in the device.The plunger was then pushed back into the handle/device.The plunger was then depressed, and a cuff miss was noted when the plunger was removed (step3).The suture of one new prostyle device was successfully pre-placed.The sheath was upsized to a 13f, and the ablation procedure was completed.Hemostasis was achieved with the pre-placed prostyle suture.There was no reported adverse patient effect.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: medical device problem code: 2017 - use after damage.
|
| |
|