MAUDE Adverse Event Report: ABBOTT VASCULAR INC. Perclose¿ ProStyle¿; Device, Hemostasis, Vascular

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. Perclose¿ ProStyle¿; Device, Hemostasis, Vascular Back to Search Results
Model Number 12773-03
Medical Device Problem Codes Failure to Cycle (1142); Unintended System Motion (1430); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this was venous closure of the right common femoral vein using a prostyle device using the pre-close technique via a 10f sheath hole prior to an ablation interventional procedure.Reportedly, during step 2, the plunger was pulled back/dislodged (protruding) but still in the device.The plunger was then pushed back into the handle/device.The plunger was then depressed, and a cuff miss was noted when the plunger was removed (step3).The suture of one new prostyle device was successfully pre-placed.The sheath was upsized to a 13f, and the ablation procedure was completed.Hemostasis was achieved with the pre-placed prostyle suture.There was no reported adverse patient effect.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.H6: medical device problem code: 2017 - use after damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Perclose¿ ProStyle¿
Common Device Name
Device, Hemostasis, Vascular
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
ABBOTT VASCULAR, Reg # 3005718570 (P099)
cashel road
clonmel tipperary UNK
EI   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50869951
Report Number2024168-2026-03665
Device Sequence Number10715164
Product Code MGB
UDI-Device Identifier08717648235184
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P960043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number12773-03
Device Catalogue Number12773-03
Device Lot Number6052741
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/08/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/27/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-