Beckman Coulter, Inc Access 2 Module DxC 500i; Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
|
Back to Search Results |
|
| Model Number |
Access 2 Module DxC 500i |
| Medical Device Problem Code |
No Device Output (1435)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, the customer reported that their dxc 500i access 2 instrument (part number: c13252; serial number (b)(6)) was generating a timeout error and causing the instrument to remain in a busy state without processing a sample.The customer reported that one troponin patient sample experienced a turnaround delay, however there was no report of any impact to patient treatment.There was no injury, death, or delay/change in treatment related to the reported issue and there was no report of harm for a patient, an operator or any other person.
|
| |
|
Additional Manufacturer Narrative
|
|
A1: the full patient identifier is (b)(6).A2, a3, a4, a5 and a6: the customer did not supply patient demographics such as age, date of birth, sex, weight, ethnicity or race.H3 and h6:
the customer informed the cts agent that their dxc 500i access 2 instrument was generating a timeout error and causing the instrument to remain in a busy state without processing a sample.The customer indicated that the error was related to the access side, specified by the event code ¿0x7c0028¿, indicating a processing timeout caused a rack to be moved from the immunoassay analyzer sample aspiration lane to the rack buffer area.The customer stated that they had already resolved the issue by performing a complete system shutdown and reboot but wanted the issue to be documented.In conclusion, the cause behind the timeout error and the instrument remaining in a busy state without processing a sample is unknown per the available information.The customer resolved the issue after performing a complete system shutdown and reboot.
|
| |
|