Model Number iS3-BR
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/02/2026
Type of Reportable Event Malfunction
Event or Problem Description
Through follow-up, the following additional information was received.Per report, the surgeon confirmed that the trocar was initially observed to be bent while inside the anterior chamber, and was subsequently removed from the eye without being activated.Reportedly, the surgeon touched the bent trocar with forceps outside of the eye, and the trocar then broke with no possibility of any fragment remaining in the eye.
 
Additional Manufacturer Narrative
Additional information: b5, g2, g3.The additional information received through follow-up was reassessed for reportability criteria; and concluded that the reported event no longer meets the statutory definition of a medical device reportable event.Based on the clarification received, glaukos no longer considers this report as a reportable event.Mfr# reference:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
iStent Infinite Trabecular Micro-Bypass System
Common Device Name
Aqueous Shunt
Manufacturer (Section D)
Glaukos Corporation
229 avenida fabricante
san clemente CA 92672
Manufacturer (Section G)
Glaukos Corporation
229 avenida fabricante
san clemente CA 92672
Manufacturer Contact
brittany massey
229 avenida fabricante
san clemente, CA 92672
+1(949)367-9600
MDR Report Key50870021
Report Number2032546-2026-00086
Device Sequence Number16117566
Product Code KYF
UDI-Device Identifier00853704002470
UDI-Public0100853704002470103311971727093021331197BR0061
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K220032
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberiS3-BR
Device Lot Number331197
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Supplement Date Received by Manufacturer09/23/2026
Initial Report FDA Received Date09/23/2026
Supplement Report FDA Received Date10/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
Patient RaceWhite
-
-