The customer reported receiving adult mobile aneroids that hole in the cuff and would not inflate during use.Medline received two adult mobile aneroids (bladder and inflation bulb only) in used condition, attached a control gauge to the cuffs, tested the samples according to the instructions for use, and tried to inflate each bladder.The complaint was confirmed; medline was unsuccessful in inflating each bladder, confirmed that they had an air leak, and upon closer visual inspection noticed the samples had a tear along the seam on the opposite side of the tubing.The cause was cannot determine cause with received sample; per vendor, through internal investigation including dhr review and inspection records review, no abnormalities were found and the complaint could not be confirmed, and the division will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|