| Model Number |
977119 |
| Medical Device Problem Codes |
High impedance (1291); Malposition of Device (2616); Charging Problem (2892)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Fluid Discharge (2686)
|
| Date of Event |
09/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from a patient regarding an implantable neurostimulator (ins).The reason for the call was patient said they are having an issue with charging the ins.Patient stated that it is very hard to charge and it is very painful.When it finally makes a connection, it is like it is sticking up on one end like it's ready to pop through the skin.If they touch it just right with the charger it is an excruciating pain, otherwise it is just very terrible.The issue started on (b)(6) 2026 where the ins was oozing and that wednesday they started to have pain while charging while on vacation.Patient has an appointment with their healthcare provider (hcp) on thursday.Patient stated their hcp said to call and coordinate for a local representative to be present.The patient was redirected to their hcp to further address the issue.
|
| |
|
Event or Problem Description
|
|
Additional information was received from the manufacturer representative (rep).It was reported that there were high impedances.Electrode 1-32547.2-40000.3-26249.Reprogramming.During planned pocket revision on 9/25/26 the surgeon will check leads and impedances.A device site infection was also reported with fluid discharge and pain at the ins site.Surgeon has planned to revise ins in the pocket and check for infection.The issue is ongoing.The manufacturer representative (rep) indicated they would send any further information as they become aware.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|