The account reported finding dressing not sticking; the complaint also states the suresite dressing was not adhering properly and had peeled off of patients.A sample was received for the investigation; 3 ea were received and 20 ea were pulled from inventory, and the samples returned and the samples from inventory had visual presence of the adhesive and passed functional testing by application and removal from a test surface.Finished good production records and finished good pack build were reviewed, with no run without approvals, deviations or non-conformance reports and no issues with the build that likely contributed to the reported issue.No defects were identified; per the complaint details and sample received, the reported issue was unable to be confirmed, the root cause cannot be determined, and this issue will be monitored by centurions complaint system.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|