Catalog Number MPH4540
Medical Device Problem Codes Computer Software Problem (1112); Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "err4 before even counting down and it definitely has enough blood and testing solution." "it throws an error at four before it even starts counting down.".
 
Additional Manufacturer Narrative
The customer reported receiving a medline evencare proview glucose meter that was giving an error message during use; there was no reported death or serious injury associated with this event.Medline received the meter in used condition with working batteries and tested it using control evencare proview test strips and control solutions; the meter turned on and the blood icon appeared, but an "er4" message was displayed during testing for a total of three times, and visual inspection confirmed the presence of an adequate amount of control solution.The complaint was confirmed by medline; the vendor reported that, through internal investigation including dhr review and inspection records review, no abnormalities were found and the complaint could not be confirmed.No root cause was identified in the provided investigation, and medline concluded that the meter did not work as intended; no trends were identified for the specified material number and corresponding defect type.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
METER,GLUCOSE,EVENCARE PROVIEW,PROF.USE
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
ayan asim
three lakes drive
northfield, IL 60093--275
MDR Report Key50870182
Report Number1417592-2026-49020
Device Sequence Number9224079
Product Code CGA
UDI-Device Identifier10888277409163
UDI-Public10888277409163
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K111456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMPH4540
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/06/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2025
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-