| Catalog Number |
03039773190 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 1 patient sample tested with cholesterol gen.2 assay on a cobas 4000 c311 stand alone system.Initial result: 418.4 mg/dl.The customer noticed that the initial result was high.No questionable result was reported outside the laboratory, and the sample was repeated.Repeat result: 218.5 mg/dl.The repeat result was deemed to be correct.
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas 4000 c311 stand alone system serial number was (b)(6).The qc was acceptable.The investigation is ongoing.
|
| |
|