• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Roche Diagnostics Cholesterol Gen.2; enzymatic esterase--oxidase, cholesterol

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Roche Diagnostics Cholesterol Gen.2; enzymatic esterase--oxidase, cholesterol Back to Search Results
Catalog Number 03039773190
Medical Device Problem Code Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
We received an allegation about discrepant results for 1 patient sample tested with cholesterol gen.2 assay on a cobas 4000 c311 stand alone system.Initial result: 418.4 mg/dl.The customer noticed that the initial result was high.No questionable result was reported outside the laboratory, and the sample was repeated.Repeat result: 218.5 mg/dl.The repeat result was deemed to be correct.
 
Additional Manufacturer Narrative
The cobas 4000 c311 stand alone system serial number was (b)(6).The qc was acceptable.The investigation is ongoing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Cholesterol Gen.2
Common Device Name
enzymatic esterase--oxidase, cholesterol
Manufacturer (Section D)
Roche Diagnostics
9115 hague road
indianapolis IN 46250 -045
Manufacturer (Section G)
Roche Diagnostics GmbH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
natalie catton
9115 hague road
indianapolis, IN 46250
MDR Report Key50870258
Report Number1823260-2026-03907
Device Sequence Number18820401
Product Code CHH
UDI-Device Identifier04015630918430
UDI-Public04015630918430
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K031824
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03039773190
Device Lot Number93818201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-