| Catalog Number |
04745914001 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of questionable tina-quant albumin gen.2 - urine application results for 1 patient on a cobas 6000 c501 module.The initial result was 52.5 mg/l.The repeat result was 7.7 mg/l.Additional repeat results were 31.6 mg/l and 7.5 mg/l.The results on a cobas c311 on (b)(6) 2026 were 10.8 mg/l and 5.8 mg/l.The results on the c501 on (b)(6) 2026 were 6.8 mg/l and 7.1 mg/l.A precision test was performed with the original sample, yielding the following results:
25.9 mg/l,
44.8 mg/l,
41.0 mg/l,
17.2 mg/l,
13.9 mg/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The tina-quant albumin gen.2 - urine application reagent lot number is 926603, and the expiration date was not provided.A field service engineer (fse) replaced the rinse unit tubing, reagent nozzles, and a reagent valve.The fse thoroughly checked and cleaned all fluidic and vacuum lines,¿ including the tank.Qc recovery was within specifications after these actions.The investigation is ongoing.
|
| |
|