A complaint was received, and the customer reported receiving a medline forehead strip thermometer that is off by a full degree; there was no reported death or serious injury associated with this event.Medline received a 100 pack of mdsts101 and two photos from the customer; the photographs showed the strips being compared while not being on the center of the forehead, and testing resulted in 96.8f for both control and test strips sent in by customer, with an oral thermometer result of 97.5f.Per the instructions for use, these readings are within the 1.0f accuracy, and the complaint is not confirmed.Based on the review of the details provided for the complaint investigation, medline was unable to confirm the complaint issue or that the product failed to meet specifications.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|