| Model Number |
CONSOLE PLUS |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report that essenz venous clamp (vc) was not working properly, even if it was calibrated.As troubleshooting, the tube was removed and then placed in again but the problem was not solved.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A1-a5 patient information was not provided.H11 livanova field service engineer was disptached to customer facility to inspect the device.Log files were reviewed and no errors were found.Functional verification test was performend with no issue and the device returned to service in its expected functions.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|