The customer reported that the product was not holding a charge/won't charge and that the new battery pack will not charge, voltage reading is 0.Medline received 1 each lift part, charger, in used condition; upon opening the product, everything was intact, the cord was intact, the charger was tested in a known working volt meter and it was 27.53, and the charger's light did turn green when plugged into a known working outlet.The investigation results were not confirmed, and the customer complaint has been not confirmed.No root cause was identified; the division will monitor through trending, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|