MAUDE Adverse Event Report: Terumo Medical Corporation ANGIO-SEAL; DEVICE, HEMOSTASIS, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Terumo Medical Corporation ANGIO-SEAL; DEVICE, HEMOSTASIS, VASCULAR Back to Search Results
Model Number 610131
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
Terumo received the following reported information: the doctor did the location of blood with dilator and put the device in.When he went to pull back nothing hooked or came out.The device slid out of the arteriotomy.When inspected the account stated the footplate and collagen were still in the sheath.There was no harm to the patient; however, manual pressure was held.The procedure performed was a carotid angiogram.No blood loss was reported.
 
Additional Manufacturer Narrative
A1: patient identifier: requested, not provided a2: age & date of birth: requested, not provided a3a: sex: requested, not provided a4: weight: requested, not provided a5: ethnicity: requested, not provided a6: race: requested, not provided d6a: implanted date: device was not implanted d6b: explanted date: device was not explanted e3: occupation: inventory manager neuro - rn the actual device has been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIO-SEAL
Common Device Name
DEVICE, HEMOSTASIS, VASCULAR
Manufacturer (Section D)
Terumo Medical Corporation
265 davidson ave
suite 320
somerset NJ 08873
Manufacturer (Section G)
Terumo Puerto Rico LLC
innovacion street lot 21
caguas west industrial park
caguas 00725
PR   00725
Manufacturer Contact
stephanie handy
950 elkton blvd
elkton, MD 21921
+1(949)989-0491
MDR Report Key50870951
Report Number3013394970-2026-00553
Device Sequence Number13089930
Product Code MGB
UDI-Device Identifier00389701011813
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P930038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number610131
Device Catalogue Number610131
Device Lot Number0001652232
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/11/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/30/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-