AIZU OLYMPUS CO., LTD. GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-Q150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the gastrointestinal videoscope had a separate issue.During the initial evaluation of the device, it was observed that the specified resolving power was not obtained due to damage on charged coupled device unit.No patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the final investigation of device and the initial evaluation findings, the specified resolving power was not obtained due to damage on charged coupled device unit was confirmed.A root cause could not be identified.Based on the results of the investigation, the most probable cause of the malfunction is traced to component failure.Date of event is unknown.Additional information will be provided if available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50870953
Report Number9610595-2026-78321
Device Sequence Number15781519
Product Code FDS
UDI-Device Identifier04953170307485
UDI-Public04953170307485
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K051645
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberGIF-Q150
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/26/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-