| Model Number |
HexaVue SCC 1.0 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that the hexavue integration system for their hexavue custom room rack was experiencing intermittent image capture issues.This event did not occur during a procedure.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite after the reported event to inspect the unit and found the cxps frame needed to be reset.To resolve the issue, the technician reset the cxps frame.The technician tested the unit, confirmed it to be operating according to specification, and returned it to service.The technician instructed the user facility on how to reset the cxps frame.No additional issues have been reported.
|
| |
|