• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; NEEDLE,HYPODERM,SAFETY, 25GX1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; NEEDLE,HYPODERM,SAFETY, 25GX1 Back to Search Results
Catalog Number SSN100255
Medical Device Problem Codes Disconnection (1171); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the nurses have been complaining about the needle on the syringe coming off, not staying on while giving vaccine, they stated it is sliding and they have to hold the actual needle when injecting vaccine into arm.
 
Additional Manufacturer Narrative
The customer reported multiple incidents of the needle "loosening prior to administering" when attaching the needle to "pre-filled influenza vaccine" syringes, and reported that this issue led to the needle "popping off" and "propelling through the air" after administering the vaccination.No samples or photos were provided; investigation included review of the lot/serial number and recent complaint trends/scar activity for the past 12 months, and per vendor, through sample investigation and dhr review, no abnormalities were found and the product quality meets the requirements of product specifications.The complaint was not confirmed.Per vendor, the safety injection needles produced by caina have no quality issues, and the vendor inferred that the needle detachment was caused by the customer not tightening the safety injection needle and syringe during use; based on review, medline was unable to confirm the complaint issue or that the product failed to meet specifications, and there was no reported death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
NEEDLE,HYPODERM,SAFETY, 25GX1
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
sullivan bluhm
three lakes drive
northfield, IL 60093--275
MDR Report Key50871133
Report Number1417592-2026-49153
Device Sequence Number9380299
Product Code FMF
UDI-Device Identifier30193489031984
UDI-Public30193489031984
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K213054
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberSSN100255
Device Lot Number89724050001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2024
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-