MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
| Model Number |
8637-20 |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received from a patient regarding an external device to an implantable pump infusing unknown morphine for non-malignant pain.It was reported that the patient stated that it took a long time to connect the handset to the pump because it was getting stuck at 90%.An agent assisted the patient with deleting the data, and after that, the patient was able to connect to the pump without any lag.However, the patient remained adamant about replacing the handset as the healthcare provider (hcp) suspected that the handset might have an issue.The agent reviewed the information and sent a repair request to replace the handset.Additional information was later received from the patient, who confirmed they had been experiencing the 90% lag.The patient did not have the handset with them to conduct the troubleshooting the agent offered.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Date is approximate.Month and year are confirmed valid.D10.Section d information references the main component of the system.Other relevant device(s) are: product id: a820, product type: software product id: hh90t01, serial# (b)(6), product type: mobile platform medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|