• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical MitraClip G5 Delivery System; mitral valve repair devices Back to Search Results
Model Number CDS0801-XTW
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/11/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported on (b)(6) 2026 that the patient presented with grade 4+ degenerative mitral regurgitation (mr), pre-existing chordal rupture, restricted and flail posterior leaflet for a mitraclip procedure.During the procedure, first mitraclip xtw implanted on anterior -posterior commissure.Post deployment, posterior leaflet was observed to be ruptured.Second mitraclip xtw was implanted lateral to first mitraclip.The first mitraclip xtw had single leaflet device attachment (slda) from the posterior leaflet due to interaction between two clips.The mr was reduced to grade 3+ post procedure.There was no clinically significant delay reported.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints from the lot.All information was investigated and based on the information provided, the reported device damaged by another device (caused damage) appears to be due to procedural conditions associated with positioning and manipulation.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MitraClip G5 Delivery System
Common Device Name
mitral valve repair devices
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
Abbott Medical Northpoint Reg 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50871218
Report Number2135147-2026-06427
Device Sequence Number9223610
Product Code NKM
UDI-Device Identifier05415067050779
UDI-Public(01)05415067050779(17)271105(10)60508A1030
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-XTW
Device Catalogue NumberCDS0801-XTW
Device Lot Number60508A1030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/08/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Steerable Guide Catheter
Patient Age78 YR
Patient SexMale
Patient Weight138 LB
Patient RaceWhite
-
-