| Model Number |
CDS0801-XTW |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 that the patient presented with grade 4+ degenerative mitral regurgitation (mr), pre-existing chordal rupture, restricted and flail posterior leaflet for a mitraclip procedure.During the procedure, first mitraclip xtw implanted on anterior -posterior commissure.Post deployment, posterior leaflet was observed to be ruptured.Second mitraclip xtw was implanted lateral to first mitraclip.The first mitraclip xtw had single leaflet device attachment (slda) from the posterior leaflet due to interaction between two clips.The mr was reduced to grade 3+ post procedure.There was no clinically significant delay reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no other complaints from the lot.All information was investigated and based on the information provided, the reported device damaged by another device (caused damage) appears to be due to procedural conditions associated with positioning and manipulation.There is no indication of a product issue with respect to manufacture, design or labeling.
|
| |
|