| Model Number |
D152 |
| Medical Device Problem Codes |
Failure to Convert Rhythm (1540); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient with this implantable cardioverter defibrillator (icd) had three separate ventricular tachycardia (vt) episodes and therapy was exhausted in two of the episodes.Additionally, this icd exhibited undersensing.Technical services (ts) reviewed the data and noted therapy was appropriate and that there was some rare undersensing.Ts noted the undersensing did not impact the time to delivery therapy.No adverse patient effects were reported.This icd remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|