| Model Number |
D233 |
| Medical Device Problem Codes |
Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
10/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) delivered an inappropriate shock due to a supraventricular tachycardia (svt).Technical services (ts) recommended reprogramming to change from tachy zone to a shock box so they can treat the patient with medicines or ablation.Additional information received approximately four years later reported that this icd delivered three bursts of anti-tachycardia pacing (atp) and a single shock due to svt.The shock broke the svt and normal sinus rhythm resumed.Technical services (ts) discussed troubleshooting options and programming considerations.The icd remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.The product was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
|
| |
|