• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the cystonephrofiberscope had a separate issue.During the initial evaluation of the device it was observed, that the bending section rubber cover had blown off and was missing.No patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to failure of the bending section cover.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Common Device Name
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
+1(508)207-7661
MDR Report Key50871585
Report Number3002808148-2026-33511
Device Sequence Number18827144
Product Code FAJ
UDI-Device Identifier04953170411144
UDI-Public04953170411144
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221690
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCYF-5
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/26/2026
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-