• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. Sphere-9; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. Sphere-9; PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH Back to Search Results
Model Number AFR-00001
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a cardiac ablation procedure for atrial fibrillation and atypical atrial flutter, ablation included the left and right pulmonary veins, posterior wall, and an anterior mitral line, with 108 pulsed field lesions and 7 radiofrequency lesions.One pulsed field lesion temperature was above 46 °c, and one radiofrequency lesion temperature was above 74 °c.A therapeutic activated clotting time of 350 was not achieved despite additional heparin boluses.Activated clotting time values were 264 at insertion, 547, 282, 453, and 282 during follow-up, and 282 at the end of the procedure. during removal of the catheter, congealed gel-like material was observed adhered to the catheter tip.The irregularly shaped pink and yellow material originated from the left atrium. the outcome of this case is unknown.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Sphere-9
Common Device Name
PERCUTANEOUS CARDIAC ABLATION CATHETER FOR TREATMENT OF ATRIAL FIBRILLATION WITH
Manufacturer (Section D)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer (Section G)
MEDTRONIC, INC.
880 technology park drive
billerica MA 01821
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
+353(091)708096
MDR Report Key50871748
Report Number3012520654-2026-00521
Device Sequence Number10715217
Product Code QZI
UDI-Device Identifier00763000993269
UDI-Public(01)00763000993269(17)270906(10)0234357104
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P240013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAFR-00001
Device Catalogue NumberAFR-00001
Device Lot Number0234357104
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured08/07/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-