| Model Number |
A22041A |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the resection sheath to the service center for repair replacement noted to have a fault tip cracked.During inspection of the device, the ceramic insulation was found to be chipped, broken off.There was no patient involvement in this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation confirmed the reported issue.Based on the investigation results, a definitive root cause cannot be identified.Probable causes include part damage from impact, stress, or excessive force.Olympus will continue to monitor the field performance of this device.
|
| |
|