• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO Manufacturing USA, LLC COMPACT INTUITIV; Unit, Phacofragmentation

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO Manufacturing USA, LLC COMPACT INTUITIV; Unit, Phacofragmentation Back to Search Results
Model Number SCP680300
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the system shut down towards the end of the case, during irrigation/aspiration (ia).No patient injury was reported.No additional information was provided.
 
Additional Manufacturer Narrative
Section a2 a3, a4 and a5: unknown/ not provided section d6a: if implanted, give date: not applicable, as the device is not an implantable.Section d6b: if explanted, give date: not applicable, as the device is not an implantable.Device evaluation: a field service engineer (fse) visited site and ran system in irrigation/aspiration (ia) mode for 1 hour and system did not shutdown.Could not duplicate problem.System is performing to specifications.A review of records including device history record and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPACT INTUITIV
Common Device Name
Unit, Phacofragmentation
Manufacturer (Section D)
AMO Manufacturing USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
+1(714)247-8552
MDR Report Key50871809
Report Number3012236936-2026-000279
Device Sequence Number18806885
Product Code HQC
UDI-Device Identifier05050474573628
UDI-Public(01)05050474573628
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K133115
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSCP680300
Device Catalogue NumberSCP680300
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-