Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative:
h3, h6: investigation summary
the product was returned to depuy synthes for evaluation, and an investigation was conducted in accordance with depuy synthes procedures.Visual inspection of the returned device found that the threaded tip of the extractor handle was broken, the broken fragment remained inside on the threaded t-handle.The possible root cause of failure can be attributed to a combination of potential factors: a high user input load and partially seated threaded connection interface, not all of which are fully understood at this time, which resulted in loading conditions that exceeded the local shear strength of the material.This condition contributed to the thread deformation and/or fracture events.However, without concrete evidence or further information, it is not possible to determine a root cause.The reported complaint was confirmed based on the evaluation results.The investigation traced the potential cause to cause not established.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history review
the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|