DePuy Ireland - 9616671 ATTUNE; Attune Instruments : Insertion Devices Back to Search Results
Catalog Number 254401005
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a total knee arthroplasty surgery, the device had an unspecified malfunction.During evaluation, it was determined that the wing knob screw component on the attune femoral introducer device was broken.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: h3, h6: the product was returned to depuy synthes for evaluation.Visual analysis of the returned attune femoral introducer found that the wing knob screw component was broken.The broken fragment was not returned for evaluation.Additionally, the sample exhibits signs of multiple use for a long period of time.The potential cause is associated with high cycle fatigue of the ml slide screw on either side of the cross pin, where the cross-sectional area is the smallest.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific instructions for use for information related to end of life, reprocessing instructions, and inspection procedures.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the attune femoral introducer would have contributed to a device issue.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE
Common Device Name
Attune Instruments : Insertion Devices
Manufacturer (Section D)
DePuy Ireland - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50871915
Report Number1818910-2026-16445
Device Sequence Number9425546
Product Code HWR
UDI-Device Identifier10603295130215
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
Exempt
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number254401005
Device Lot NumberAB3782277
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-