| Model Number |
977119 |
| Medical Device Problem Codes |
Therapy Delivered to Incorrect Body Area (1508); Battery Problem (2885); Charging Problem (2892)
|
| Health Effect - Clinical Codes |
Pain (1994); Inappropriate Device Stimulation of Tissue (4928)
|
| Date of Event |
09/18/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implantable neurostimulator remained implanted and in service, and the patient reported pain at the implantable neurostimulator site.After recharging to 40%, the charge reportedly lasted approximately 2 days.The battery reportedly depleted to 0% and the neurostimulator shut off on two occasions.The patient began charging the device daily when it reached 40%.While lying down, the patient reported that stimulation vibrated into the stomach, which had not been previously experienced.Telephone troubleshooting was performed, and an in-person appointment was scheduled to interrogate the battery.The issue remained ongoing and unresolved.No adverse outcome was reported.The manufacturer representative (rep) indicated they would send any further information as they become aware.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|