ICU Medical Costa Rica Ltd. Primary PLUM Set, CLAVE Secondary Port, CLAVE Y-Site, Secure Lock, 103 Inch; Set, administration, intravascular
|
Back to Search Results |
|
| Catalog Number |
14687-28 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch reported that when rn was back priming secondary and noted leaking at the cassette.Rn stopped back priming and clamped secondary line due to dripping.Upon further inspection, it was found that there was a crack in the connector from the secondary to the cassette under the blue piece.The rn opened the chamber of the pump, and the connector broke off from the cassette, staying attached to the secondary line.The event occurred during priming.
|
| |
|
Additional Manufacturer Narrative
|
|
Received one (1) used.List #14687-28.Primary plum set and one (1) used.List #unknown.Chemolock connector with one (1) used.List #unknown.Chemolock connector and one (1) used.List #unknown.Extension set with drip chamber and one (1) used.List #unknown.Chemolock connector for evaluation.As received the secondary port clave was separated and broken off of the cassette.A dhr lot # review could not be conducted because no lot numbers were identified.A rigid and uneven break was observed on the cassette and the clave port, indicative of a bending force.The complaint of cracked clave can be confirmed.The probable cause is due to an unintentional bending force.
|
| |
|