| Catalog Number |
151106010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patients presents in the er with pain.Exam suggests movement of the static spacer utilizing lps components.In the or, there is movement between the taper from the femoral sleeve adapter and the lps segment.Existing modular components were removed.Tapers were thoroughly cleaned and dried and new components were placed successfully with no delay nor patient harm.No visible damage was found.Doi-(b)(6) 2026.Dor-(b)(6) 2026.Affected side-right.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Investigation summary: according to the information provided, it was reported that patients presents in the er with pain.Exam suggests movement of the static spacer utilizing lps components.In the or, there is movement between the taper from the femoral sleeve adapter and the lps segment.Existing modular components were removed.Tapers were thoroughly cleaned and dried and new components were placed successfully with no delay nor patient harm.This was an unusual occurrence in the surgeon's experience as the patient routinely utilizes these components in this fashion and had very rarely has seen this complication.No visible damage was found.Doi-(b)(6) 2026.Dor-(b)(6) 2026.Affected side-right.Since the complaint device was not returned , we cannot determine a root cause for the reported failure.If the device or additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added : d10 concomitant.
|
| |
|
Search Alerts/Recalls
|