| Model Number |
MC172150SX-AL |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when the microcatheter headway-17 was opened the device was straight, but when the microguide was advanced by 0.014 mm, the doctor realized it wasn't progressing.He tried repeatedly, but it was never possible.The device was inspected and found to be fractured in the proximal portion.Another device is used to resolve the case.The patient is in good clinical condition.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was stated to be available for return to the manufacturer for evaluation, but has not yet been returned despite attempts made.Additional due diligence attempts are ongoing.The alleged product issue/event as described could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.
|
| |
|