Catalog Number 151106010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 09/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patients presents in the emergency room with pain.Exam suggests movement of the static spacer utilizing lps components.In the operating room, there is movement between the taper from the femoral sleeve adapter and the lps segment.Existing modular components were removed.Tapers were thoroughly cleaned and dried and new components were placed successfully with no delay nor patient harm.No visible damage was found.Doi-(b)(6) 2026.Dor-(b)(6) 2026.Affected side- right.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Investigation summar: according to the information provided, it was reported that patients presents in the er with pain.Exam suggests movement of the static spacer utilizing lps components.In the or, there is movement between the taper from the femoral sleeve adapter and the lps segment.Existing modular components were removed.Tapers were thoroughly cleaned and dried and new components were placed successfully with no delay nor patient harm.This was an unusual occurrence in the surgeon's experience as the patient routinely utilizes these components in this fashion and had very rarely has seen this complication.No visible damage was found.Doi-(b)(6) 2026.Dor-(b)(6) 2026.Affected side-right.Since the complaint device was not returned , we cannot determine a root cause for the reported failure.If the device or additional information is received in the future, we will reopen the complaint and perform the investigation as appropriate.It has been determined that no corrective or preventive actions are proposed.There is no indication that a design or manufacturing issue caused the reported condition.As part of depuy synthes quality processes, all devices are manufactured, inspected, and released in accordance with approved specifications.Additional monitoring for potential safety signals is conducted through complaint trend analysis as part of post-market safety surveillance activities.Device history lot: there is no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.Added : d10 concomitant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE
Common Device Name
Knee Femoral Adaptor
Manufacturer (Section D)
DePuy Ireland - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthpaedic dr.
warsaw, IN 46581
+1(908)386-3534
MDR Report Key50872040
Report Number1818910-2026-16450
Device Sequence Number18820412
Product Code MBH
UDI-Device Identifier10603295503415
UDI-Public(01)10603295503415(17)310228(10)JC8970
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K253197
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number151106010
Device Lot NumberJC8970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/03/2026
Supplement Date Received by Manufacturer09/25/2026
Initial Report FDA Received Date09/23/2026
Supplement Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATT REV SLV LPS FEM ADP +10; ATT REV SLV LPS FEM ADP +10; LPS SEGMENTAL COMPONENT 25MM; LPS SEGMENTAL COMPONENT 40MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
Patient SexMale
Patient Weight224 LB
-
-