It was reported that there may be a hole and it does not stay inflated.Medline received one medline handheld aneroid in used condition, tested the sample according to the instructions for use, attempted to inflate the cuff, and then removed the bulb and used a control inflation bulb to test the sample again.The complaint was confirmed; with the control inflation bulb, the cuff inflated up to 300 mmhg and no issues were observed during testing.The issue was caused due to a defective inflation bulb, as a result of the air mechanism being broken/blocked, preventing it from properly inflating/deflating the cuff, and the sample did not work as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|