| Model Number |
480422-03 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/07/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the 8mm vessel sealer extend instrument had a different lot number.There was no report of patient involvement or patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The vessel sealer extend (vse) instrument was analyzed and found to have missing retaining ring from one of the inputs, resulting in adjacent damage and a dislodged bearing and pitch input disk, which prevented the instrument from being placed on an in-house system for functional testing.The complaint was not confirmed by failure analysis.
|
| |
|