Abbott Medical (CRM-Sylmar) AVEIR AR 2 Leadless Pacemaker Back to Search Results
Model Number LSP203A
Medical Device Problem Codes Separation Failure (2547); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the patient presented for implant procedure.During surgery, the atrial leadless pacemaker (lp) was unable to be redocked to the delivery catheter after fixation.The tethers were cut on the delivery catheter and a single snare was used to retrieve the lp.Outside the body, the tether was unable to be removed from the docking button on the lp.A different lp was used to complete the procedure.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR AR 2 Leadless Pacemaker
Common Device Name
Leadless Pacemaker
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50872134
Report Number2017865-2026-20214
Device Sequence Number18486695
Product Code PNJ
UDI-Device Identifier05415067051189
UDI-Public(01)05415067051189(10)S000114153(17)280528
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P150035
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSP203A
Device Catalogue NumberLSP203A
Device Lot NumberS000114153
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/28/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Delivery Catheter
Patient SexUnknown
-
-