The complaint was received related to a "ripped" sensicare pi green surgical glove; the customer reported, "we had another glove rip," and there was no reported death or serious injury associated with this event.Medline observed tearing on the returned glove, performed trending over the last 2 years of complaints, confirmed the lot inspection was completed with passing results, and the vendor's review of batch records and retention samples found no abnormalities or deviations.Based on the sample returned, medline was able to confirm the complaint issue or that the product failed to meet specifications; the cause was recorded as "cannot determine cause w/received sample." no trending was identified at this time, the investigation indicated no need for further action, and complaints are monitored via the trend analysis procedure.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|