It was reported that during a da vinci-assisted radical transvesical surgical procedure, when the gas hose was connected, the pressure blew out the inlet pin of the universal seal.The customer then switched to another seal, and which blew out again.The customer used another universal seal and proceeded with the case.The procedure was completed.No known impact or patient consequence was reported.Intuitive surgical, inc.(isi) followed up with the initial reporter and obtained the following additional information: during the procedure, a broken luer connector caused a slow, partial loss of insufflation with no error messages displayed on the insufflator.The issue was resolved intraoperatively when the doctor closed the small valve and switched to a replacement seal, which restored normal conditions.The only clinical impact noted was temporarily dim visualization, which returned to normal once the new seal was connected.There were no patient injuries, and no additional medical intervention was required.The procedure was completed robotically without further complications.
|