The customer reported that the issue occurred during use, no injury occurred, and the account reported finding ra separated from manifold.A photo was received to aid our investigation; no physical sample was available, and finished good production records and finished good pack build were reviewed with no non-conformances noted that would have contributed to the reported issue.The close-up photo depicted the end of the manifold, with the ra separated from it, no other components were visible, confirming the customer's report.Without the physical sample available for review, it is impossible to determine if there was a successful swage of the ra to the manifold stem, therefore, we are not able to identify the root cause at this time; the division will monitor this issue for trending purposes.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|