Catalog Number 709-04-58F
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 08/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Left total hip revision.Revising the cup because it had spun out and broken all of the screws.We removed the cup and replaced it with a larger multihole cup and screws with mdm liner.No further information from the facility.
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT II TRITANIUM MULTI 58F
Common Device Name
Prosthesis, hip, semi-constrained, metal/polymer, cemented
Manufacturer (Section D)
Stryker Orthopaedics-Mahwah
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
Stryker Orthopaedics-Cork
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
+1(201)831-5000
MDR Report Key50873006
Report Number0002249697-2026-00876
Device Sequence Number9223661
Product Code JDI
UDI-Device Identifier07613327344059
UDI-Public07613327344059
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K171768
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number709-04-58F
Device Lot Number97575301A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-