The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the cause of the reported leak and thrombus were unable to be determined.The reported patient effect of thrombus, as listed in the triclip g5 system instructions for use, is a known possible complication associated with triclip procedures.The reported medication and medical intervention were a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|