According to the customer, the gloves were not completely leak-proof, which meant that the surgeons had bloody fingers after the operations; the reported concern was tearing during use.No photo or sample was received, and inspection records for the reported sku lot were reviewed and the lot passed with no abnormalities noted; scar and capa records for the glove division were reviewed and there have been no opened scars or capas for the reported sku in the last 24 months.In the absence of a physical sample or detailed photos, medline was unable to confirm the presence of a defect during this investigation, and the true root cause could not be accurately determined.There were no reports of death or serious injury associated with this event, and medline's quality team will continue to monitor for trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|