| Model Number |
CDS0801-XTW |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on (b)(6) 2026 that the patient presented with grade 4+ functional mitral regurgitation (mr) for a mitraclip procedure.During the procedure, after the grasping the leaflets the mitraclip xtw was unable to open.Troubleshooting was performed and was unsuccessful.As a result, the clip was implanted in it's existing position.Additional mitraclip was implanted to treat the residual jet.The mr was reduced to grade 1-2 post procedure.There was no clinically significant delay or adverse patient effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported inability to open the clip.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
| |
|