| Model Number |
4471 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker was alerted that a lead safety switch (lss) from bipolar to unipolar configurations was triggered due to high out of range pacing impedance measurements on the right ventricular (rv) lead were recorded.The health care professional (hcp) called technical services (ts) for further review of daily threshold and impedance measurements trend report.Upon review, the device data showed rv lead pacing impedances were stable within normal limits except for two instances that caused the lss.The presenting electrogram (egm) showed no noise events and isometrics was unable to reproduce the high out of range pacing impedances.Ts discussed possible causes and provided programming recommendations.No adverse patient effects were reported.At this time, this rv lead and pacemaker remain in service.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|