BOSTON SCIENTIFIC CORPORATION FINELINE II EZ Sterox; Implantable Lead Back to Search Results
Model Number 4471
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this pacemaker was alerted that a lead safety switch (lss) from bipolar to unipolar configurations was triggered due to high out of range pacing impedance measurements on the right ventricular (rv) lead were recorded.The health care professional (hcp) called technical services (ts) for further review of daily threshold and impedance measurements trend report.Upon review, the device data showed rv lead pacing impedances were stable within normal limits except for two instances that caused the lss.The presenting electrogram (egm) showed no noise events and isometrics was unable to reproduce the high out of range pacing impedances.Ts discussed possible causes and provided programming recommendations.No adverse patient effects were reported.At this time, this rv lead and pacemaker remain in service.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINELINE II EZ Sterox
Common Device Name
Implantable Lead
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
mark mischke
4100 hamline avenue north
saint paul, MN 55112
+1(612)875-7234
MDR Report Key50873392
Report Number2124215-2026-50758
Device Sequence Number13447557
Product Code DTB
UDI-Device Identifier00802526265006
UDI-Public(01)00802526265006(17)081120
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P960004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/20/2008
Device Model Number4471
Device Catalogue Number4471
Device Lot Number445555
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2026
Initial Report FDA Received Date09/23/2026
Date Device Manufactured11/20/2006
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age68 YR
Patient SexMale
-
-